IEC 62304 is the international standard for the software life cycle of medical devices. It tells you how to plan, build, test, release and maintain software so that the evidence of safe development exists before anyone asks for it. If your product is a medical device, or runs inside one, it is the document your quality system, your FDA submission and your investors' diligence team will all measure your engineering against. The practical answer for a healthtech company is to treat it as a clinical and risk question first and an engineering question second, because the safety class you assign decides how much work follows.
That is where a physician-led view helps. A safety class is a judgment about how software can hurt a patient. It is not a coding decision, and it needs someone who has seen care delivered to make it.
TL;DR
- IEC 62304 defines life cycle processes for medical device software, covering both software that is a device and software embedded in a device.
- Your software safety class (A, B or C) is set by how the software could contribute to a hazardous situation. That class sets how much documentation and verification you owe.
- A software development plan, traceability, risk-linked verification, a maintenance plan and archived release records are the core evidence.
- The standard does not cover validation or final release of the device. Those sit with system-level requirements and FDA's own expectations.
- Classify early and with clinical input. Reclassifying after the architecture is built is the most expensive way to find an error.
Table of Contents
- What IEC 62304 is and what it covers
- Why investors and founders both care
- The software safety classes, in plain terms
- What the standard asks you to produce
- How it fits with FDA requirements
- Where healthtech teams go wrong
- A sensible order of work
- Our perspective
- Working with The StartUp MD
- FAQ
- Sources
What IEC 62304 is and what it covers
IEC 62304 defines the life cycle requirements for medical device software, setting out a common framework of processes, activities and tasks. According to the FDA's recognized-standards entry, it applies to development and maintenance of software when the software is itself a medical device, or is an embedded or integral part of the final medical device. The same entry notes that it does not cover validation and final release of the device, even when the device consists entirely of software.
Two terms matter here:
- SaMD (software as a medical device): software that performs a medical purpose on its own, such as an app that analyzes images or a clinical decision support engine.
- SiMD (software in a medical device): software that runs inside or drives hardware, such as the firmware of an infusion pump.
The standard applies to both. It reaches software that runs on a processor or is executed by other software on a processor, regardless of storage or delivery method. So a cloud-hosted application and an installed one fall under the same logic.
The standard is also narrower than many founders assume. It governs the software life cycle. It does not replace your quality management system, your risk management process, or the usability work your product needs. It plugs into them.
Which edition applies in the US
The FDA recognizes IEC 62304 Edition 1.1 (2015-06), which consolidates the 2006 original with Amendment 1, as a complete standard. That is identical to ANSI AAMI IEC 62304:2006/A1:2016, per the FDA recognition entry. ISO lists the 2006 edition with its single amendment as current and confirmed in 2023. When a customer or auditor says "62304," the working text is the consolidated version with Amendment 1.
Why investors and founders both care
For a founder, conformance to a recognized standard is a practical route to showing FDA that your process is sound. For an investor, it is a signal about engineering discipline. A company that cannot show a development plan, a traceability matrix or a classification rationale is carrying hidden remediation cost, and that cost usually surfaces at the worst moment: a premarket submission, a partner security review or a Series B diligence request.
The quality regulation has moved in the same direction. FDA's revised Part 820, now called the Quality Management System Regulation, became effective on February 2, 2026. FDA inspectors now work to an updated inspection process. For software teams, the lesson is that design and development records are read as part of one connected system, not as separate piles of paper. Our guide on design controls for SaMD under 21 CFR 820.30 and IEC 62304 walks through how those records line up.
The software safety classes, in plain terms
Every software system under the standard gets a safety class. The class is the lever that sets how much rigor the rest of the standard demands.
| Class | Plain meaning | Practical effect |
|---|---|---|
| A | The software cannot contribute to a hazardous situation, or any contribution is made acceptable by risk controls outside the software | Lightest set of required activities |
| B | The software can contribute to a hazardous situation that could cause non-serious injury | Adds architecture and unit-level work |
| C | The software can contribute to a hazardous situation that could cause death or serious injury | Full set of activities and the deepest verification |
The 2006 original assigned class purely on the severity of possible harm. Amendment 1 changed the logic. You first ask whether the software can contribute to a hazardous situation at all. Then you ask whether the risk is acceptable once you account for risk control measures that sit outside the software, such as a hardware interlock or a clinician's required review step. Only if risk remains unacceptable does severity separate Class B from Class C.
A worked example
Take a triage tool that ranks incoming patient messages by urgency for a nurse team.
- Hazard: a missed high-urgency message could delay care for a deteriorating patient.
- Contribution: the ranking software plainly contributes to that hazard.
- External controls: if every message is still read by a nurse within a fixed window regardless of rank, that process is a control outside the software. If nurses work the queue in ranked order and rarely look below the top, the control is weak.
- Class: if the external control is weak and the delay could cause serious harm, you are in Class C. If a robust human check reliably catches errors, a lower class may be defensible, and you must document why.
Notice what decided the class: a clinical workflow fact, not a line of code. This is why classification rationales written by engineers alone tend to be optimistic, and why reviewers push back on them. We cover related documentation in six documents FDA reviewers want for clinical decision support.
Classify the system, then decompose with care
The class applies to the software system, and individual items inside it can be assigned lower classes only if you segregate them and can show the segregation holds. Many teams skip this and simply declare the whole system at the highest class they fear, which inflates cost. Others decompose without a defensible architecture, which collapses under review. Either way, the decision belongs in your risk file with its reasoning.
What the standard asks you to produce
IEC 62304 is organized as a set of processes. The ones founders meet first are these.
Software development planning
Per clause 5.1 of the consolidated text, the software development plan applies to all classes, A, B and C. It addresses the processes to be used, the deliverables of each activity and task, traceability between system requirements, software requirements, system testing and risk control measures, configuration and change management including SOUP, and software problem resolution.
SOUP means "software of unknown provenance": any third-party or open-source component you did not develop under your own controlled process. Most modern SaaS stacks are full of it. Each SOUP item needs to be identified and tracked in your configuration management, because a vulnerable or abandoned library becomes your problem.
Requirements, architecture and design
You derive software requirements from system requirements, and, for higher classes, document architecture and detailed design. Each requirement should trace forward to a test and back to the system need or risk control that justified it. This chain is what auditors sample.
Verification and release
Clause 5.8 requires that verification is complete and documented, that known residual anomalies are documented and, for Classes B and C, evaluated for their effect on safety, that the released version is documented, that records are archived for the longer of the device's lifetime or the regulatory retention period, and that a reliable delivery procedure exists. Release here means release of the software for use at the system level. It is not the same as device validation, which is why teams working on SaMD also look at software validation expectations and CSA.
Maintenance
Clause 6 requires a maintenance plan covering how you handle feedback, the criteria for treating a report as a problem, how risk management is used, and how you evaluate upgrades, bug fixes, patches and SOUP obsolescence. For a SaaS product that ships weekly, this clause deserves as much attention as development. A continuous delivery pipeline is compatible with the standard only when each change is assessed against risk and recorded.
Risk management ties it together
IEC 62304 expects your software activities to feed and draw on your risk management file, which most teams build to ISO 14971. Software hazards are identified, controls become requirements, and verification proves the controls work. Consistent risk discipline matters in any field with downside exposure, and risk management strategies in other domains share the same core habit: define the loss you cannot accept before you act, and track it throughout. For device software, our piece on ISO 14971 artifacts for software lists the six records teams need to be audit ready.

How it fits with FDA requirements
IEC 62304 is a standard, not a law. FDA recognizes it, and following a recognized standard is a well-trodden way to show that your process meets expectations. But a premarket submission is judged against FDA's own software guidance. FDA's Medical Device Software Guidance Navigator lists software-related guidance and states that its listings are not comprehensive of everything that may apply to a given device.
The June 2023 final guidance on the content of premarket submissions for device software functions uses two documentation levels, Enhanced and Basic. Enhanced applies where a software failure could present a hazardous situation with a probable risk of death or serious injury, assessed before risk controls are applied. Basic covers everything else. Notice that FDA assesses this before controls, while Amendment 1 of 62304 assesses after external controls. The two frameworks are related but not identical, and a team should document both views rather than assume one answer carries over.
A few adjacent topics usually come up alongside 62304:
- IEC 82304-1 covers system-level requirements for standalone health software, including validation and final release, which 62304 leaves out. The CSA adoption notes that system-level activities for standalone software devices belong there.
- Usability engineering under IEC 62366 produces evidence reviewers expect. See our IEC 62366 software evidence plan.
- Cybersecurity is a separate premarket expectation. Our three cybersecurity deliverables and seven-item checklist cover it, and the SBOM-related regulatory strategy piece connects it to your SOUP inventory.
- Pathway choice between De Novo and 510(k) shapes how much evidence you assemble and when.
- Whether you are a device at all is a real question for some products. See our steps on MDDS and avoiding a 510(k).
Where healthtech teams go wrong
The same patterns repeat across early-stage companies.
Classifying by engineering instinct. Without clinical context, teams pick Class A because the software "only displays" information. A display that a clinician acts on can still contribute to a hazard.
Writing documents after the build. Retroactive traceability is slow and error-prone. Planning first costs less than reconstructing intent from commit history.
Ignoring SOUP. A modern product may depend on hundreds of packages. Without an inventory and a policy for updates, maintenance records are empty.
Treating AI components as exempt. Machine learning models still run as software, so they sit inside the life cycle. Their data, training and change control need to be planned, and their failure modes need to be in the risk file.

Letting the standard sit with one engineer. Safety class decisions affect product scope, go-to-market timing and valuation. Founders, clinical leadership and quality need to own them together.
A sensible order of work
- Confirm the product's regulatory status. Decide whether it is SaMD, SiMD or outside device regulation. Everything else depends on this.
- Define intended use and users in clinical terms. A precise intended use makes hazards identifiable.
- Run a hazard analysis with clinical input. List how the software could contribute to harm, and which controls exist outside the software.
- Assign the safety class and write the rationale. Record the reasoning, not just the letter.
- Write the software development plan. Include traceability, configuration management, SOUP and problem resolution.
- Build and verify against the plan. Keep verification evidence linked to requirements and risk controls.
- Plan release and maintenance. Archive records, define how field feedback is triaged, and decide how patches are assessed.
- Review before submission or diligence. Walk the file as a reviewer would, from a hazard through to a passing test.
The heaviest clinical judgment sits in steps 2 to 4. That is where an experienced physician leader saves the most time, since those decisions are hard to reverse.
Our perspective
Most 62304 trouble I see starts before a line of documentation is written. A company decides what its software does for a patient, and nobody with clinical responsibility tests that statement against how care is actually delivered.
The safety class is a clinical judgment dressed in engineering language. It asks what happens to a patient when the software is wrong, and what else in the workflow would catch the error. A physician who has worked at the bedside and built a business reads that question differently from a team that has only done one of the two.
That is the gap a fractional Chief Medical Officer is built to close. Our approach is hands-on: direct access to an advisor with both clinical and business expertise, who helps connect intended use, hazard analysis, regulatory strategy and investor expectations into one coherent story.
If your classification rationale, your development plan or your diligence materials feel misaligned with how your product is really used, I'd welcome a conversation.
Paul Bergeron MD, MBA
Working with The StartUp MD
The StartUp MD is a consultancy for healthcare SaaS and healthtech companies, led by Paul Bergeron, MD, MBA, who brings more than 25 years of experience in medicine and business. We offer Fractional Chief Medical Officer leadership and advisory services that include clinical strategy, product evaluation, regulatory guidance and investor support. For founders, that means a clinical leader in the room when software is classified and risks are weighed. For investors, it means a physician's read on portfolio company readiness.

You can see how the engagements work on our services page, or start from thestartupmd.com.
FAQ
Is IEC 62304 mandatory?
The standard itself is voluntary, but FDA recognizes it, and conforming to a recognized standard is a common way to demonstrate that your software process is sound. In practice, customers, auditors and investors expect it for device software. Your regulatory obligations come from FDA requirements, and the standard is how many teams meet the software portion.
Does IEC 62304 apply to SaaS and cloud software?
Yes, when the software is a medical device or an integral part of one. The standard applies to software that runs on a processor regardless of how it is stored or delivered. A hosted clinical decision support tool with a medical purpose is within its scope. Software with no medical purpose generally is not.
What is the difference between IEC 62304 and IEC 82304-1?
IEC 62304 covers the software life cycle: planning, development, verification, release for use at the system level, and maintenance. IEC 82304-1 addresses system-level requirements for standalone health software, including validation and final release. Standalone SaMD teams typically need to consider both.
Who should decide the software safety class?
The decision should combine engineering, quality and clinical judgment. Class depends on how software can contribute to a hazardous situation and what controls exist outside the software, which are clinical workflow questions. Clinical leadership, such as a fractional Chief Medical Officer, helps make the rationale defensible.